CategoryParameterSpecification
Measurement Principle & PerformanceMeasurement PrincipleTPPS4 spectrophotometry
Measurement Range0–0.2 mg/L;0–1 mg/L;0–2 mg/L
Limit of Detection (LOD)≤0.005 mg/L
Indication ErrorAt concentrations ≤0.04 mg/L: ±0.005 mg/L
At concentrations ≥0.04 mg/L: ±10%
Repeatability≤5%
Zero Drift±5%
Span Drift±10%
Measurement Time<50 minutes
Measurement & Quality ControlMeasurement ModesInterval measurement (user-selectable from 0–9,999 minutes); scheduled measurement at selected hours over a 24-hour cycle; single measurement; externally triggered measurement
Calibration & Quality ControlManual calibration; scheduled automatic calibration; automatic verification using standard solutions at user-defined times
Communication & SignalsCommunication InterfacesRS-232;RS-485;RJ45
Analog Signal4–20 mA
Digital Inputs / Outputs2 digital outputs; 2 digital inputs
Operation & Data ManagementRegulatory Platform IntegrationTFT full-color LCD touchscreen for display and command input
Display & ControlsDisplays measurement, idle, fault, and maintenance states; provides fault, overrange, limit-exceedance, and low-reagent alarms
Status Display & AlarmsRecords instrument operations; stores at least one year of raw measurement data and operating logs
Data Logging & StorageAutomatically saves data during a power failure; resumes operation after a fault reset or power restoration
Power-Loss Protection & Automatic RecoveryThree-level user access control
Installation & Operating ConditionsUser Access ControlRecommended ambient temperature: 5–40°C; relative humidity: ≤85%, non-condensing
Environmental Requirements220 ±22 V AC;50 ±0.5 Hz
Power Supply≤150 W
Power Consumption600 × 450 × 1,555 mm
Copy linkWeChatFacebookLinkedInWhatsAppPinterest