CategoryParameterSpecification
Measurement Principle & PerformanceMeasurement PrincipleAmmonium molybdate spectrophotometry
Measurement Range0–2 mg/L; 0–10 mg/L (extendable)
Limit of Quantification (LOQ)≤0.020 mg/L (indication error: ±30%)
Indication ErrorAt concentrations <0.4 mg/L: ±0.04 mg/L
At concentrations ≥0.4 mg/L: ±10%
Repeatability≤2%
24-hour Low-Concentration Drift≤0.01 mg/L
24-hour High-Concentration Drift≤1%
Measurement Time<45 minutes
Measurement & Quality ControlMeasurement ModesInterval measurement (user-selectable from 0–9,999 minutes); scheduled measurement at selected hours over a 24-hour cycle; single measurement; externally triggered measurement
Calibration & Quality ControlManual calibration; scheduled automatic calibration; automatic verification using standard solutions at user-defined times
Communication & SignalsCommunication InterfacesRS-232;RS-485;RJ45
Analog Signal4–20 mA
Digital Inputs / Outputs2 digital outputs; 2 digital inputs
Operation & Data ManagementRegulatory Platform IntegrationTFT full-color LCD touchscreen for display and command input
Display & ControlsDisplays measurement, idle, fault, and maintenance states; provides fault, overrange, limit-exceedance, and low-reagent alarms
Status Display & AlarmsRecords instrument operations; stores at least one year of raw measurement data and operating logs
Data Logging & StorageAutomatically saves data during a power failure; resumes operation after a fault reset or power restoration
Power-Loss Protection & Automatic RecoveryThree-level user access control
Installation & Operating ConditionsUser Access ControlRecommended ambient temperature: 5–40°C; relative humidity: ≤85%, non-condensing
Environmental Requirements220 ±22 V AC;50 ±0.5 Hz
Power Supply<150 W
Power ConsumptionCompact enclosure: 374 × 320 × 783 mm; large enclosure: 600 × 450 × 1,555 mm